FT-111, User Manual Rev. 2.0.0, June 2022 Page 64 of 137
[6-- CALB / ADJUST] Calibration and Adjustment Block
[
6--
CALIB / ADJUST
]
SCALE ZERO AND SPAN SETTING MAIN BLOCK
Press key sequentially to access this main block,
or press or key to enter configuration parameters,
or press key to go to the next block,
or press key to exit from programming.
[
61-
CALIBRATION
]
CALIBRATION
Press key or key again to enter this menu.
Or press key to go to the next sub-block.
[
611
TEST WEIGHT
]
M
Scale calibration with test weight
[
612
LINEARIZATIO
]
M
Multipoint scale calibration to increase linearization
[
613
ELECTRONIC
]
M
eCal electronic calibration without test weight
[
62-
ADJUSTMENT
]
ADJUSTMENTS
Press key or key again to enter this menu.
Or press key to go to the next sub-block.
[
621
ZERO ADJUSTM
]
M
Zero adjustment
[
622
SPAN ADJUSTM
]
M
Span adjustment
[
623
S UNDER LOAD
]
M
Span adjustment of loaded scale
Span adjustment under load
Span adjustment with temporary zeroing
[
624
GRAVITY CAL
]
M
Gravity acceleration of the place of the calibration.
[
625
GRAVITY USAG
]
M
Gravity acceleration of the place of the usage.
[
63-
COEFFICIENTS
]
CALIBRATION COEFICIENTS
Press key or key again to enter this menu.
Press key to go to beginning of the sub-block or press key to
go to the next main-block.
[
631
LOAD COEFFIC
]
M
The load weight used at the calibration is indicated here.
[
632
ZERO COEFFIC
]
M
This coefficient is determined the zero point of the scale.
[
633
GAIN COEFFIC
]
M
This coefficient is related with the gain factor of the scale.