FT-111, User Manual Rev. 2.0.0, June 2022 Page 63 of 137
[
523
MAX
:
]
M
Scale capacity and division (d)
CAP1/d1
CAP2/d2
CAP3/d3
Enter scale capacity and division after press
key. Capacities and divisions of MR and MI scales are
entered as Cap1, d1, Cap2, d2, Cap3, d3.
[
524
OVER
:
9d
]
M
Limit of Indication
NO
1d
9d
2%
5%
:
:
:
:
:
Over indication after Max
1 division more than Max
9 divisions more than Max
2% more than Max
5% more than Max
[
52A
BATTERY
:
NO
]
M
Battery management.
NO
12V
BPAC
:
:
:
Disables battery management.
Enables battery management for 12V battery.
Enables battery management for FT-Accu-Ex2 battery
[
52B
AUTO OFF
:
5
]
M
Auto power off time for battery management.
NO
1
:
:
Disabled Auto Power OFF.
The scale is not used along 1 minute.
2
3
4
5
10
:
:
:
:
:
The scale is not used along 2 minutes.
The scale is not used along 3 minutes.
The scale is not used along 4 minutes.
The scale is not used along 5 minutes.
The scale is not used along 10 minutes.
53-
DLC SETUP
]
DIGITAL LOAD CELL SETUP (Only FT-111D)
Press key or key again to enter this menu.
Or press key to go to the next sub-block.
[
532
QUANTITY
:
01
]
M
Quantity of digital load cell
Page 72
Enter the quantity of DLC used in the scale.
The quantity can be entered between 1 and 16.
[
533
ADDRESSING
]
M
Addressing of digital load cells
Enter serial number after press key.
[
54-
SHIFT ADJUST
]
SHIFT / ECCENTRICITY ADJUSTMENT (Only FT-111D)
Press key or key again to enter this menu.
Press key to go to the beginning of the sub-block or press
key to go to the next main-block.
[
541
METHOD
:
CELL
]
M
Adjustment method
Page 73
CELL
PAIR
:
:
Individual load cell shift adjust
Sectional pair shift adjustment
[
542
AUTO ADJUST
]
M
Automatic Eccentricity Adjustment
The eccentricity correction of scale is performed
automatically.
[
543
MANUAL ADJUS
]
M
Manual Adjustment
The eccentricity correction of scale is performed manually
[
544
SET TO 1
:
NO
]
M
Temporarily set shift constants to 1
NO
YES
:
:
Normal operation
(entered shift constants are used).
Test mode (shift constants are equal to 1).