FT-111, User Manual Rev. 2.0.0, June 2022 Page 61 of 137
[5-- SCALE] Scale Block
The most important parameter is 511 which limits the use of the parameters in main blocks 5, 6 and 8. The
selections of this parameter are
Selections of parameters at set up and calibration are free for
industrial usage of the instrument.
Industrial weighing
only gross
Autozeroing, taring and power on zero are disabled.
(Recommended for tank /silo weighing in gross.)
Industrial weighing net
(Taring can be enabled)
Autozeroing and power on zero are disabled.
(Recommended for tank /silo weighing in gross.)
Approved scale according to
OIML
Metrology related parameters are restricted to limits of OIML R76
and EU type approval of the instrument.
After setting parameter 511, even you set any parameter out of the accepted range of selection of par 511, it
is saved in the acceptable limit. For example, if taring is activated at INDG selection, it will be disabled while
exiting from set up. For approved scales, set the parameters in main blocks 5, 6 and 8 perform calibration
carefully due to sealing of the scale in legal usage.
SCALE RELATED PARAMETERS MAIN BLOCK
Press key sequentially to access this main block,
or press or key to enter configuration parameters,
or press key to go to the next block,
or press key to exit from programming.
SCALE SET UP
Press key or key again to enter this menu.
Or press key to go to the next sub-block.
Industrial. All parameters can be freely selected.
Industrial weighing of tank, hopper or silo in gross.
(Taring, AZTrack and Power on Zero are disabled)
Industrial weighing in Net of tank, hopper or silo
(disabled AZT and Power on Zero, enabled Tare)
OIML approved scale.
Temporary indication with key
Toggle. Start and end high resolution by pressing key in
sequence.
Always high resolution
Disable.
± 2%
± 2%, [ POWER ON ZERO ERR ] prompt cannot be
erased. Call service.
± %10
+ %15, - %5
± %20