FT-111, User Manual Rev. 2.0.0, June 2022 Page 74 of 137
Setting Shift Coefficients of all Load cells to 1
Setting all coefficients to 1 temporary might be needed to service the scale without losing the shift
coefficients.
1. Press key at the [ 544 SET TO 1 : NO ] prompt to adjust the parameter.
2. Parameter changes to [ 544 SET TO 1 : YES ], press key until [ SAVE : YES ] prompt appears
on the display.
3. Press key to save the changes.
Do not forget to reload coefficients after testing the scale as;
1. Press key at the [ 544 SET TO 1 : YES ] prompt to adjust the parameter.
2. Parameter changes to [ 544 SET TO 1 : NO ], press key until [ SAVE : YES ] prompt appears
on the display.
3. Press key to save the changes into the memory.