Issue/Rev. 0.2 (12/12)
Page 116 • MN0M023
8 Block INPUTS Device Functions Proline Promass 83
VALUE 20 mA
(5203)
Use this function to assign a value to the 20 mA current.
User input:
5-digit floating-point number
Factory setting:
Dependent on the process variable assigned to the current input (see function ASSIGN
CURRENT INPUT, 5200).
– Pressure: 100 barg
– Temperature: 200 °C
– Reference density: 2.00 kg/Nl
! Note!
The appropriate unit is taken from the functions UNIT PRESSURE (0426), UNIT
TEMPERATURE (0422) or UNIT REFERENCE DENSITY (0421).
ERROR VALUE
(5204)
Use this function to enter a defined error value for the process variable concerned.
If the current value lies outside of the selected range (see function CURRENT SPAN,
5201), then the process variable is set to the “error value” defined here and a
corresponding notice message CURRENT INPUT RANGE (# 363) is generated.
User input:
5-digit floating-point number
Factory setting:
Dependent on the process variable assigned to the current input (see function ASSIGN
CURRENT INPUT, 5200).
– Pressure: 50 barg
– Temperature: 75 °C
– Reference density: 1.25 kg/Nl
! Note!
• Triggered amplifier faults or the error behavior of the outputs do not have any
influence on the current input.
• The appropriate unit is taken from the functions UNIT PRESSURE (0426), UNIT
TEMPERATURE (0422) or UNIT REFERENCE DENSITY (0421).
Function description
INPUTS o CURRENT INPUT o CONFIGURATION