Device Functions Proline Promass 83 6 Block TOTALIZERS
6.1 Group TOTALIZER (1...3)
6.1.1 Function group CONFIGURATION
TOTALIZERS D
TOTALIZER 1 DAA
CONFIGURATION 300
TOTALIZER 2 DAB
CONFIGURATION 300
TOTALIZER 3 DAC
CONFIGURATION 300
Function description
TOTALIZERS o TOTALIZER (1...3) o CONFIGURATION
The function descriptions below apply to totalizers 1...3; the totalizers are independently configurable.
ASSIGN
(3000)
Use this function to assign a measured variable to the totalizer in question.
Options (standard):
OFF
MASS FLOW
VOLUME FLOW
CORRECTED VOLUME FLOW
Advanced options with optional software package CONCENTRATION:
TARGET MASS
TARGET VOLUME
TARGET CORRECTED VOLUME
CARRIER MASS
CARRIER VOLUME
CARRIER CORRECTED VOLUME
Factory setting:
MASS FLOW
! Note!
• The totalizer is reset to “0” as soon as the selection is changed.
• If you select OFF in the function group CONFIGURATION of the totalizer in question,
only the ASSIGN (3000) function remains visible.