Device Functions Proline Promass 83 11 Block SUPERVISION
11.1.2 Function group OPERATION
SUPERVISION J
SYSTEM JAA
CONFIGURATION 800
OPERATION 804
Function description
SUPERVISION o SYSTEM o OPERATION
ACTUAL SYSTEM
CONDITION
(8040)
Use this function to check the present system condition.
Display:
“SYSTEM OK” or the fault / notice message with the highest priority.
PREVIOUS SYSTEM
CONDITION
(8041)
Use this function to view the fifteen most recent fault and notice messages since
measuring last started.
Display:
The 15 most recent fault or notice messages.
SIMULATION FAILSAFE
MODE
(8042)
Use this function to set all inputs, outputs and totalizers to their defined failsafe modes, in
order to check whether they respond correctly. During this time, the message
“SIMULATION FAILSAFE MODE” appears on the display.
Options:
ON
OFF
Factory setting:
OFF
SIMULATION
MEASURAND
(8043)
Use this function to set all inputs, outputs and totalizers to their defined flow-response
modes, in order to check whether they respond correctly. During this time, the message
“SIMULATION MEASURAND” appears on the display.
Options:
OFF
MASS FLOW
VOLUME FLOW
CORRECTED VOLUME FLOW
DENSITY
REFERENCE DENSITY
TEMPERATURE
Factory setting:
OFF
" Caution!
The measuring device cannot be used for measuring while this simulation is in
progress.
The setting is not saved if the power supply fails.