Issue/Rev. 0.2 (12/12)
Page 26MN0M023
4 Block QUICK SETUP Device Functions Proline Promass 83
QUICK SETUP
BATCHING/DOSING
(1005)
! Note!
This function is only available when the optional software package BATCHING is
installed.
Use this function to start the (optional) application specific Setup menu for batching.
Options:
YES
NO
Factory setting:
NO
! Note!
You will find a flowchart of the BATCHING setup menu on page 32. For more
information on Setup menus, please refer to the Operating Instructions
Proline Promass 83, BA 059D/06/en.
T-DAT SAVE/LOAD
(1009)
Use this function to save the parameter settings / configuration of the transmitter in a
transmitter DAT (T-DAT), or to load the parameter settings from the T-DAT into the
EEPROM (manual safety function).
Application examples:
After commissioning, the current measuring point parameters can be saved to the
T-DAT as a backup.
If the transmitter is replaced for some reason, the data from the T-DAT can be loaded
into the new transmitter (EEPROM).
Options:
CANCEL
SAVE (from EEPROM to T-DAT)
LOAD (from the T-DAT into EEPROM)
Factory setting:
CANCEL
! Note!
If the target device has an older software version, the message “TRANSM. SW–DAT”
is displayed during startup. Then only the “SAVE” function is available.
LOAD
This function is only possible if the target device has the same software version as, or a
more recent software version than, the source device.
•SAVE
This function is always available.
Function description
QUICK SETUP