3 Block MEASURED VARIABLES Device Functions Proline Promass 83
TEXT ARBITRARY
CONCENTRATION
(0606)
! Note!
This function is not available unless the optional software package CONCENTRATION is
installed.
Use this function to enter a text for the selectable concentration unit
(user-defined density unit).
User input:
xxxxxxx (max. 4 characters)
Valid characters are A-Z, 0-9, +,
-, decimal point, white space or underscore
Factory setting:
“_ _ _ _” (No text)
Example:
Enter text “HFCS” (for High Fructose Corn Syrup).
FACTOR ARBITRARY
CONCENTRATION
(0607)
! Note!
This function is not available unless the optional software package CONCENTRATION is
installed and an option selected in the function DENSITY FUNCTION (7000),
(see page 141).
Use this function to define a factor for the selectable concentration unit, see function
TEXT ARBITRARY CONCENTRATION (0606).
User input:
7-digit floating-point number
Factory setting:
1
Reference quantity:
Arbitrary concentration unit / %
Example:
The measured concentration 1% should be output as 0.01 HFCS value User input o
0.01 [HFCS]
Function description
MEASURED VARIABLES o SPECIAL-UNITS o ARBITRARY UNIT