Issue/Rev. 0.2 (12/12) MN0M023 • Page 99
Device Functions Proline Promass 83 7 Block OUTPUTS
Function description
OUTPUTS o PULSE/FREQUENCY OUTPUT (1...2) o OPERATION (STATUS)
ACTUAL STATUS
(4341)
! Note!
This function is not available unless the STATUS setting was selected in the OPERATION
MODE function (4200).
Use this function to check the current status of the status output.
Display:
NOT CONDUCTIVE
CONDUCTIVE
SIMULATION SWITCH
POINT
(4342)
! Note!
This function is not available unless the STATUS setting was selected in the OPERATION
MODE function (4200).
Use this function to activate simulation of the status output.
Options:
OFF
ON
Factory setting:
OFF
! Note!
The “SIMULATION STATUS OUTPUT” message indicates that simulation is active.
The measuring device continues to measure while simulation is in progress, i.e. the
current measured values are output correctly via the other outputs.
" Caution!
The setting is not saved if the power supply fails.
VALUE SIMULATION
SWITCH POINT
(4343)
! Note!
This function is not available unless STATUS was selected in the OPERATION MODE
function (4200) and the SIMULATION SWITCH POINT function (4342) is active
(= ON).
Use this function to define the switching response of the status output during the
simulation. This value is used to test downstream devices and the measuring device itself.
Options:
NOT CONDUCTIVE
CONDUCTIVE
Factory setting:
NOT CONDUCTIVE
" Caution!
The setting is not saved if the power supply fails.