9 Block BASIC FUNCTION Device Functions Proline Promass 83
EPD EXC. CURR. MAX
(6426)
Use this function to activate the empty pipe detection (EPD).
In the event of inhomogeneous fluids or air bubbles, the exciting current of the
measuring pipes increases. If the exciting current set in this function is exceeded, error
message #700 is output similar to the functions “EPD VALUE LOW (6423)” and “EPD
VALUE HIGH (6423).”
User input:
5-digit floating-point number
Factory setting:
100 mA (deactivated)
! Note!
The function is not activated until a value under 100 mA is input. Entering the value
100 mA deactivates the function.
Function description
BASIC FUNCTION o PROCESS PARAMETER o EPD PARAMETER