9 Block BASIC FUNCTION Device Functions Proline Promass 83
DENSITY ADJUST MODE
(6482)
Use this function to select whether a 1-point or a 2-point density adjustment should be
carried out.
Options:
CANCEL
1-POINT
2-POINT
DENSITY SETPOINT 1
(6483)
Use this function to enter the density setpoint value for the first fluid for which you want
to carry out field density adjustment.
User input:
5-digit floating-point number, incl. units
! Note!
• The preset density entered here should not vary from the actual fluid density by a more
than ±10%.
• The appropriate unit is taken from the function group SYSTEM UNITS (see page 17).
MEASURE FLUID 1
(6484)
In this function the actual density of the first fluid is measured for density adjustment.
Options:
CANCEL
START
DENSITY SETPOINT 2
(6485)
Use this function to enter the density setpoint value for the second fluid for which you
want to carry out field density adjustment.
User input:
5-digit floating-point number, incl. units
! Note!
• The preset density entered here should not vary from the actual fluid density by a more
than ±10%.
• The difference between the density setpoint values must be at least 0.2 kg/l.
• The appropriate unit is taken from the function group SYSTEM UNITS (see page 17).
MEASURE FLUID 2
(6486)
In this function the current density of the second fluid is measured for density
adjustment.
Options:
CANCEL
START
Function description
BASIC FUNCTION o PROCESS PARAMETER o ADJUSTMENT