Issue/Rev. 0.2 (12/12) MN0M023 • Page 171
Device Functions Proline Promass 83 10 Block SPECIAL FUNCTION
10.3.6 Function group TEMPERATURE
SPECIAL FUNCTION H
ADVANCED DIAGNOSIS HEA
CONFIGURATION 740
ACQUISITION 741
MASS FLOW 742
DENSITY 743
REFERENCE DENSITY 744
TEMPERATURE 745
Function description
SPECIAL FUNCTION o ADVANCED DIAGNOSIS o TEMPERATURE
! Note!
The appropriate unit is taken from the function UNIT TEMPERATURE (0422) (see page 21).
REFERENCE VALUE
TEMPERATURE
(7450)
The reference value for the temperature appears on the display.
Display:
5-digit floating-point number, including unit and sign
TEMPERATURE
(7451)
The currently measured temperature appears on the display.
Display:
5-digit floating-point number, including unit and sign
MINIMUM VALUE
(7452)
The lowest temperature value since the saved values were last reset appears on the
display.
Display:
5-digit floating-point number, including unit and sign
MAXIMUM VALUE
(7453)
The highest temperature value since the saved values were last reset appears on the
display.
Display:
5-digit floating-point number, including unit and sign
TEMPERATURE HISTORY
(7454)
The last ten temperature values since the saved values were last reset appear on the
display.
Display:
5-digit floating-point number, including unit and sign
TEMPERATURE
DEVIATION
(7455)
This function displays the deviation between the currently measured temperature and the
reference values (FACTORY or USER), see page 166, selected in the function SELECT
REFERENCE CONDITION (7402).
Display:
5-digit floating-point number, including unit and sign
WARNING LEVEL
(7456)
! Note!
This function is not available unless ON was selected in the WARNING MODE function
(7403).
Use this function to specify a limit value for the temperature. A notice message is
generated if the limit value is exceeded.
User input:
0...99999 °C
Factory setting:
100 °C