11 Block SUPERVISION Device Functions Proline Promass 83
VALUE SIMULATION
MEASURAND
(8044)
! Note!
The function is not visible unless the function SIMULATION MEASURAND (8043) is
active.
Use this function to define a freely selectable value (e.g. 12 m
3
/s). This is used to test the
associated functions in the device itself and downstream signal loops.
User input:
5-digit floating-point number [unit]
Factory setting:
0 [unit]
" Caution!
The setting is not saved if the power supply fails.
The appropriate unit is taken from the function group SYSTEM UNITS (ACA),
(see page 17).
SYSTEM RESET
(8046)
Use this function to perform a reset of the measuring system.
Options:
NO
RESTART SYSTEM (restart without interrupting power supply)
Factory setting:
NO
OPERATION HOURS
(8048)
The hours of operation of the device appear on the display.
Display:
Depends on the number of hours of operation elapsed:
Hours of operation < 10 hours
o display format = 0:00:00 (hr:min:sec)
Hours of operation 10...10,000 hours
o display format = 0000:00 (hr:min)
Hours of operation > 10,000 hours o display format = 000000 (hr)
Function description
SUPERVISION o SYSTEM o OPERATION