EasyManua.ls Logo

FMC Technologies Proline Promass 83 E - Function Group TUBE DAMPING

FMC Technologies Proline Promass 83 E
206 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Issue/Rev. 0.2 (12/12)
Page 172MN0M023
10 Block SPECIAL FUNCTION Device Functions Proline Promass 83
10.3.7 Function group TUBE DAMPING
SPECIAL FUNCTION H
ADVANCED DIAGNOSIS HEA
CONFIGURATION 740
ACQUISITION 741
MASS FLOW 742
DENSITY 743
REFERENCE DENSITY 744
TEMPERATURE 745
TUBE DAMPING 746
Function description
SPECIAL FUNCTION o ADVANCED DIAGNOSIS o TUBE DAMPING
REFERENCE VALUE
TUBE DAMPING
(7460)
The reference value for tube damping appears on the display.
Display:
5-digit floating-point number
TUBE DAMPING
(7461)
The measured tube damping appears on the display.
Display:
5-digit floating-point number
MINIMUM VALUE
(7462)
The lowest tube damping value since the saved values were last reset appears on the
display.
Display:
5-digit floating-point number
MAXIMUM VALUE
(7463)
The highest tube damping value since the saved values were last reset appears on the
display.
Display:
5-digit floating-point number
TUBE DAMPING
HISTORY
(7464)
The last ten tube damping values since the saved values were last reset appears on the
display.
Display:
5-digit floating-point number
TUBE DAMPING
DEVIATION
(7465)
This function displays the deviation between the measured tube damping and the
reference values (FACTORY or USER), see page 166, selected in the function SELECT
REFERENCE CONDITION (7402).
Display:
5-digit floating-point number
WARNING LEVEL
(7466)
! Note!
This function is not available unless ON was selected in the WARNING MODE function
(7403).
Use this function to specify a limit value for tube damping. A notice message is generated
if the limit value is exceeded.
User input:
0...99999 [%]
Factory setting:
1000%

Table of Contents

Related product manuals