Issue/Rev. 0.2 (12/12)
Page 148MN0M023
10 Block SPECIAL FUNCTION Device Functions Proline Promass 83
ASSIGN BATCH
VARIABLE
(7202)
Use this function to assign a batching variable to the batching specification.
Options:
OFF
MASS FLOW
VOLUME FLOW
CORRECTED VOLUME FLOW
Advanced options with optional software package CONCENTRATION:
TARGET MASS
TARGET VOLUME
TARGET CORRECTED VOLUME
CARRIER MASS
CARRIER VOLUME
CARRIER CORRECTED VOLUME
Factory setting:
OFF
! Note!
The possible assignments of the display functions are automatically extended. Once a
batching variable has been selected (MASS or VOLUME), you can locally define the
application-specific function of the minus key (start-stop-continue) and the plus key
(stop-batching name/quantity) in the information line by means of the “batching
menu” assignment. In this way, a direct batching control station is made available
locally at the measuring device by means of the user interface and the controls.
Select OFF if the BATCHING functionality is no longer to be used. All settings related
to the function (e.g. switching contact assigned to the relay output) must be assigned
to another functionality.
BATCH QUANTITY
(7203)
Use this function to define the quantity to be batched.
User input:
5-digit floating-point number
Factory setting:
0 [unit]
! Note!
The appropriate unit is taken from the function group SYSTEM UNITS (ACA),
(see page 17).
When the batching quantity entered here is achieved, valve 1 closes (see function
CLOSE VALVE 1 (7221) on page 153).
Function description
SPECIAL FUNCTION o BATCHING FUNCTION o CONFIGURATION