GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A015 - Lift Movement Troubleshooting
Page no. 1045 Chapter 9
JC-TSG-A-015.fm
ILLUSTRATION 5 - TROUBLESHOOTING FLOWCHART FOR LIFT MOVEMENT PROBLEMS RELATED TO
ERRORS 102/013 AND 102/014 - TYPE 2 LIFT SCREW ASSEMBLY
Perform Lift Functional Checks.
See
section 3-1, Lift Functional
Checks
on page 1048.
NOK
Grease Checks
See
section 3-2, Grease
Checks
on page 1048.
Grease Lift Screw
See
section 3-3, Grease Lift
Screw
on page 1049.
NOK
OK
NOK
Lift Gas Spring Checks
See
section 3-4, Lift Gas Spring
Checks
on page 1052.
NOK
Change Lift Gas Spring
See
Job Card D/R A206A - Lift
Gas Spring
on page 1183.
Perform Lift Functional Checks.
See
section 3-1, Lift Functional
Checks
on page 1048.
OK
NOK
Perform Lift Functional Checks.
See
section 3-1, Lift Functional
Checks
on page 1048.
OK
OK
Nut, Bearing, and Lift Checks
See
section 3-5, Bearing and
Lift Checks
on page 1057.
OK
If Nut, Bearing or Lift NOK,
change for Type 2 Lift without
Motor (5391083), keeping the
motor from the old lift.
See
Job Card D/R A232 - Lift
Screw Assembly
on page 1255.
NOK
Perform Lift Functional Checks.
See
section 3-1, Lift Functional
Checks
on page 1048.
OK
Check Lift Motor
See
section 3-6, Lift Motor
Checks
on page 1057.
NOK
Change Brakeless Lift Motor
(5212900)
See
Job Card D/R A205 - Lift
Motor
on page 1179.
Perform Lift Functional Checks.
See
section 3-1, Lift Functional
Checks
on page 1048.
OK
NOK
Contact OLC.
Corrective actions
Change Lift Board
See
Job Card D/R A201 - Lift
Board
on page 1161.
NOK
Perform Lift Functional Checks.
See
section 3-1, Lift Functional
Checks
on page 1048.
OK
NOK
OK