DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
3-74 Section 3-9 - Configuration
3-9-2-5 Units of Measure
Figure 3-71 Select Units of Measure
Table 3-10 Select Units of Measure
STEP TASK EXPECTED RESULT(S)
1.
Open the Configuration Window, see Log On to the System as
ADM on page 4-10.
Select SYSTEM
, if needed.
The System Settings Window is displayed.
2.
Use the drop down Units dialog (see Figure 3-71) to select
Metric or US Units.
The selected units (Metric or US) will be used for measurements as soon
as the unit has been restarted.
3.
Restart the scanner. All the changed settings will be used after the restart.
Units