DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
3-128 Section 3-10 - Connectivity Setup
3-10-7-3 Configuring the Vivid i n/ Vivid q N for MPEGVue Export
1) Press the Config (or F2) button and log on as Adm, as described in Log On to the System as ADM
on page 4-10.
2.) Select Connectivity.
3.) Select Dataflow.
4) From the drop-down menu, select Misc Export - see Figure 3-136.
5) From selected output devices, highlight evue and then click the Properties button - refer to
Figure 3-137 on page 3-129.
Figure 3-136 Dataflow - Misc Export