GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
3-154 Section 3-10 - Connectivity Setup
3-10-9-3-4 Stopping & Running the Cine-Loop
When a loop is selected, it appears on the screen in Run-Mode.
There are several ways to stop a loop and view it frame-by-frame:
Clicking on the Freeze/Run icon (#5 in Figure 3-162) will freeze the loop. Clicking on same icon
once again will run the loop again.
Clicking on the scrolling icons (#6 in Figure 3-162) will stop the loop, allowing the user to advance
forwards or backwards to the required frame.
Using the left / right arrows - when the loop is in Run-Mode, pressing either the left or right arrows
on the PC keyboard will freeze the loop. Further pressing the left or right arrows will advance the
loop frame-by-frame.
3-10-9-3-5 Selecting a Different Patient
Click on the Back to Patient List button (#13 in Figure 3-162). On the Patient List, click on the Exam
Date of the required patient.
3-10-9-3-6 Exporting Images or Loops from MPEGvue
Using the “Save As…” function, an image may be saved to any of the following file formats:
- .BMP
- .WMV (MPEG4)
- .JPEG
To save an image, click on the diskette icon (#8 in Figure 3-162). A Save As… dialog-box will appear,
allowing you to select the desired file format in which to save the image.
3-10-9-3-7 Performing Measurements on MPEGVue Images
The following measurements can be performed on MPEGVue images:
Distance and area in 2D images
Height, time, and slope in M-mode images
Velocity and time in Doppler modes
In a 2D image:
1) Select the Measurement tool (#9 in Figure 3-162).
The distance measurement tool is automatically selected.
2) Perform the distance measurement.
3) To perform an area measurement, right-click on the image and select Area. Perform the area
measurement.
In Other Modes:
1) Select the Measurement tool (#9 in Figure 3-162).
2) Perform the measurement.
Note: If the original image contains measurements that were taken and stored on the
Vivid i n/ Vivid q N scanner, those measurements will appear in the MPEGvue image. In
this case, additional measurements cannot be performed on that image.
Note: Currently, the measurement function is not available on DICOM images/loops, any
secondary captured screens 3D images, or images with discontinuities.