DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
3-172 Section 3-10 - Connectivity Setup
5.) If required, access the System Configuration utility, as follows:
a.) Make sure the System Configuration tab is activated as shown in Figure 3-186.
System Configuration read-only information is displayed (in table format) under the following
column headings: Item, Value, Minimum Requirements.
Below this information, the eVue Operation Utility buttons enable the user to enable or
disable the eVue mode, as required. An additional button (Configure eVue) is used to perform
the eVue configuration procedure.
Note: In order to enable eVue operation, it is necessary to first perform the eVue configuration
procedure as described in the following step.
b.) Configure eVue (if required), as follows:
Click the Configure eVue button. The Configuration dialog box opens:
Figure 3-186 MPEGVue Patient Manager Window - System Configuration Tab