DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
4-4 Section 4-2 - General Procedures
Follow These Steps to Connect Mains Power to the Unit
Connecting the Vivid i n/ Vivid q N ultrasound unit involves preliminary checks of the power cord, voltage
level and compliance with electrical safety requirements.
1) Ensure that the wall outlet is of appropriate type, and that the power switch is turned OFF.
2) Uncoil the power cable, allowing sufficient slack so that the unit can be moved slightly.
3) Verify that the power cable is without any visible scratches or any sign of damage.
4) Verify that the on-site mains voltage is within the limits indicated on the rating label on the base of
the Vivid i n and Vivid q N portable ultrasound scanner.
5.) Connect the Mains Power Cable female plug via the AC Adapter to the Power Inlet (DC IN - refer
to Figure 4-1) at the rear of the unit.
6.) Connect the other end (male plug) of the Mains Power Cable to a hospital-grade mains power outlet
with the proper rated voltage.
The unit is now ready for Power ON/Boot Up.
.
Figure 4-1 Mains Power Connector - Rear Panel
DC IN