DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
4-24 Section 4-3 - Functional Checks
8
When done, press OK to close all windows.
LCD PIXEL TEST:
1
Boot-up the System in Regular Scanning Mode
2
Press Alt +Config to start the diagnostic package
3
From the diagnostic test tree, select BEP.
This will activate PC Doctor diagnostics.
4
Run System Diagnostics\Monitor Test
5
Select Solid Test. No missing pixels in R/G/B tests
6
Select Red
Make sure there are no missing pixels in the Full Screen
Red or display in another color
7
Select Green
Make sure there are no missing pixels in the Full Screen
Green or display in another color
8
Select Blue
Make sure there are no missing pixels in the Full Screen
Blue or display in another color
9 When done, press Close
The system will prompt: Did the test pass?
Respond accordingly.
EXTERNAL MONITOR TEST:
1
Connect the external monitor to the system
2
Boot-up the System in Regular Scanning Mode
3
Press Config
4
Under the Connectivity tab, select TCPIP
5
Select Advanced
6
In the address bar, type Controlbar and press Enter
7
From the Control Panel, select Display then select
Settings
8
From the Settings Tab, select Advanced. Select
S3Display tab
Make sure the CRT check box is activated.
9
When done, press OK to close all windows.
Table 4-11 Monitor Test (cont’d) - Recommended
Step Task Expected Result(s)