DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 4 - General Procedures and Functional Checks 4-25
4-3-1-8 ECG/Respiratory Test
10
Press Alt +Config to start the diagnostic package,
11
From the diagnostic test tree, select BEP.
This will activate PC Doctor diagnostics.
12
Run System Diagnostics\Monitor Test
13
Select Combination Test.
14
Follow the on-screen instructions. PASS
15
When done, press Close
The system will prompt that the test passed.
Table 4-12 ECG/Respiratory Test - Recommended
Step Task Expected Result(s)
1
Boot up the system in Maintenance Mode, using the
Service Dongle and load the application by
double-clicking the Echoloader icon on the desktop.
2
Connect the ECG cable to the input on the ECG port on
the right side of the Vivid i n/ Vivid q N scanner.
Connect a Cardiac probe (for example, 3S-RS) and
press on Physio. Set ECG as OFF and Int/Resp as ON.
Connect the ECG leads to yourself (as the simulated
patient). Refer to the instructions in the Vivid i n/Vivid q N
User Manual.
3
Press Alt + Config to enter into Diagnostics Mode
4
From the Diagnostic Test tree, select BEP.
5
Run the ECG Test
6
From the Test ECG device window, set the resolution to
1 second and press the Open Device button
7
Verify that a single clear QRS signal is displayed.
Table 4-11 Monitor Test (cont’d) - Recommended
Step Task Expected Result(s)