GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
4-32 Section 4-3 - Functional Checks
4-3-2-5 External USB Hard Disk Test
NOTE: If the External USB Hard Disk is not currently connected to the system, refer to External USB Hard Disk
Installation Procedure on page 8-151 for configuration instructions.
Table 4-19 External USB Hard Disk Test
Step Task Expected Result(s)
1
Check the External USB Hard Disk status light
indicators.
Refer to Figure 8-152 on page 8-152.
2
Create a dummy patient file on the Vivid i n or Vivid q N
scanner. Save several images in the patient file.
3
Export the images to the External USB Hard Disk.
4
Delete the dummy patient file (including images) from
the Vivid i n/ Vivid q N database.
The deleted patient file and images are no longer listed.
5
Import the dummy patient file and images from the
External USB Hard Disk.
The patient file is now listed in the Vivid i n/ Vivid q N
database.
6
Open the patient file and check that the images are
available for review.
The newly-imported images are displayed correctly on
the Vivid i n/ Vivid q N scanner.
7
Delete the dummy patient file (including images) from
the Vivid i n/ Vivid q N database.
The deleted patient file and images are no longer listed.