DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 4 - General Procedures and Functional Checks 4-31
6.) Click OK.
The system displays the Dataflow screen indicating that the device check has been successfully
performed as indicated by the green confirmation tick - see Figure 4-9.
7.) Under Selected Devices (Figure 4-9), proceed to the next device (in this example "Remote
Database" that is under "Outputs"), click Check and proceed as before.
8.) Continue this procedure until all selected devices have been verified.
Figure 4-8 Information - Check Status Box
Figure 4-9 Selected Device "Remote Database" showing Successful Check