DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
4-36 Section 4-3 - Functional Checks
4-3-4-3 Plug and Scan Test
18
Press OK to close the message and then press Done.
A progress bar and message are displayed asking you
to wait.
19
Press Alt+E and select CD/DVD Writable (H:\) from the
menu.
The message Media Ejected is displayed.
20
Remove the DVD from the drive.
21
Verify that all of the saved exams and images of the
Test Patient exist in the database.
The exams and images that were exported and
imported for the Test Patient should appear in the
database.
22
Select the Test Patient exams and press Delete.
A message is displayed prompting the user to confirm
the delete command.
23
Press OK to confirm delete (must confirm three times). The Test Patient is deleted from patient list.
Table 4-25 Plug and Scan Test - Recommended
Task Expected Result(s)
1
Press the On/Off button on the control console for no
more than 3 seconds.
The system enters Standby mode.
2
Turn OFF the system’s circuit breaker.
The system remains in Standby mode, indicated by the
blinking On/Off button LED.
3
Turn ON the system’s circuit breaker.
Press the On/Off Button.
The system returns quickly to the Power-ON state -less
than 30 seconds.
4
Turn OFF the system’s circuit breaker. Turn ON the
system’s circuit breaker after entering to Standby mode.
The system enters Standby mode and stays in this
mode. The On/Off button LED is blinking.
5
Turn OFF the system’s circuit breaker.
Press the On/Off button.
The system enters the Power-OFF state. The On/Off
button LED is OFF.
Table 4-24 DVD/CD Writer Test (cont’d) - Recommended
Step Task Expected Result(s)