GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
6-12 Section 6-2 - LCD Display Adjustments and Calibration
6-2-11-1 Performing the Battery Calibration Procedure
NOTE: It is recommended to commence this procedure towards the end of the working day (when the
battery is fully charged) then allow the procedure to be completed overnight, as outlined in
Figure 6-10.
1) Click on the battery icon at the bottom of the scanning screen
OR
from the message window shown below, click Recalibrate.
The Battery Calibration Wizard (Page 1) opens - Figure 6-11.
2.) Make sure the AC power cable is connected to the Vivid i n/ Vivid q N, then click Next to commence
the first stage of the process (Full Charge to maximum capacity).
This step will fully recharge the battery and may take up to 4 hours.
Figure 6-11 Battery Calibration Wizard - Page 1