GE HEALTHCARE
DIRECTION FQ091013, REVISION 1    VIVID I N AND VIVID Q N SERVICE MANUAL
6-14 Section 6-2 - LCD Display Adjustments and Calibration
5.) Disconnect the AC power cable from the Vivid i n/ Vivid q N, then click Next. 
The second phase of the process (Full Discharge) commences and the Battery Calibration Wizard 
(Page 3) is displayed as shown in Figure 6-14 on page 6-15.
Figure 6-13   Battery Calibration Wizard - Page 3