GE HEALTHCARE
DIRECTION FQ091013, REVISION 1    VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting  7-7
Reports Displays the specific report (or list of reports) for 
the currently-selected test sequence. 
(shows all OK and failed tests)
Certain tests will have more than one report (see 
the example in Figure 7-3 below).
Double-clicking on a currently-selected 
report displays that report in the Report 
window (lower portion of the View Log 
dialog box) - refer to Figure 7-4 below.
Reports Scroll 
Bar and Arrows
Enables scrolling through the list of Reports A test sequence may have up to four 
reports. Since only one specific test may 
have failed, the scroll arrows enable 
jumping directly from one Failed test 
report to the next (i.e., failures only). 
Alternatively, you can scroll through all the 
reports (one-by-one).
Erase Button Clears all test sequence log files for that specific 
(currently selected) test listed in the Tests 
window (passed and failed tests).
It is not recommended to clear the test 
sequence log files unless a test failed due 
to user intervention or as a result of a false 
alarm. (It is necessary to erase each log 
file separately).
Exit Button Closes the View Log dialog box Returns to the Diagnostic Test window.
Report Window Lower portion of the View Log dialog box where the 
the currently-selected report is displayed in a 
separate window (see the example in Figure 7-3). 
The displayed Average, Deviation, Max and Min 
values are dynamically updated. 
Visible only after test execution.
Report window 
Scroll Arrows
Enables scrolling through all channels. 
Displays the currently-selected channel number. 
For example, Count: 64
Close Button Closes the Report window.
Table 7-3    Vivid i n/ Vivid q N Diagnostics Test Window - View Log
Option Description Comments
Scroll between 
reports here
Jump to Failed Test Reports