DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
7-38 Section 7-4 - Performing Front End (FE) Diagnostics
As the testing sequence progresses, the progress bar will advance to reflect the test progress. The
LED on the RFI status indicator in the Data Flow map will be highlighted in the appropriate color to
indicate the current test status, as follows:
Green: All recommended tests for these components have been completed and no problems
were identified.
Red: Problems were identified in these components during the performance of the test.
Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the RFI Analog Circuits Test sequence is complete,
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-16.
8.) If necessary, use the scroll bars and/or arrows to scroll through the list of Test results.
Note: For this test, there is one report (Analog Circuits Test report) which is listed in Reports on
the right. By default, this report is displayed in the Reports window below (Figure 7-16). Use
the Report window scroll bars and/or arrows to scroll through complete report if necessary.
9) When done, trackball to the Exit button and press Set to close the View Report window.
Figure 7-16 View Test Log - RFI Analog Circuits Test