DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-65
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
Green: All recommended tests for these components have been completed and no problems
were identified.
Red: Problems were identified in these components during the performance of the test.
Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the ACW Matrix Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-38.
Note: The ACW Matrix Test test has three test reports (Signal Frequency, Signal RMS, and
Signal SNR) which are listed in the Reports window on the right. By default, the Signal
Frequency report is displayed in the Reports window below (Figure 7-38).
Figure 7-38 View Test Log - ACW Matrix Test - Signal Frequency
Scroll between
channels here
Average Result
Expected Average
Selected channel
Selected channel # (32)