7-100 Section 7-5 - Monitoring Tests
• Yellow: The tests executed so far on these components have passed, but not all 
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Fans Speed Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details 
of which tests passed or failed, as shown in the example in Figure 7-69.
Note: For this test, there is one report (Fans Status) which is listed in the Reports window on the 
right. By default, this report is displayed in the Reports window below.
7.) Check that the test passed and that result values for each fan are within the pre-defined Max and 
Min limits, as shown in Figure 7-69.
8) When done, trackball to the Exit button and press Set to close the View Report window.
9) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when 
prompted). 
• If the test passed but the original problem persists, contact your local On-Line Center for 
assistance.
• If the test fails, repeat the Fan Speed Test.
• If errors were identified but the results were very close to the upper or lower limits, perform 
calibration, as described in DC Offset Calibration on page 7 - 25, and then repeat the test.
Figure 7-69   View Test Log - Fans Speed Test