DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
7-114 Section 7-6 - Performing Back End Diagnostics on the System
9.) Verify that a single clear QRS signal is displayed on the screen, as shown in Figure 7-79.
10.)When done, trackball to the Exit button and press Set.
11.)When prompted by the following message, select Yes or No as applicable:
12) Proceed as follows:
If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
13.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.