GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-115
7-6-1-3 PC Doctor
NOTE: For a complete description of all PC Doctor tests, refer to the information in PC Doctor on page 7 - 15.
This procedures in this section are divided into two groups. Refer to the relevant procedure, as follows:
• Automatic Tests
- Full Back End Test w/o Hard Disk Scan - see below
- Full PC Doctor Test (50 min) on page 7 - 117
- CPU Tests on page 7 - 119
- HD Test on page 7 - 122
- Memory Test on page 7 - 124
- Video Test on page 7 - 126
- PCI Test on page 7 - 128
- Network Test on page 7 - 130
7-6-1-3-1 Full Back End Test w/o Hard Disk Scan
NOTE: For a description of this test, refer to Full Back End Test without Hard Disk on page 7 - 15.
1.) In the Diagnostic Test window, trackball to Full Back End Test w/o Hard Disk Scan in the list of PC
Doctor tests available for selection from the Diagnostic Test Tree (see Figure 7-80 below) and press
Set. The Test name is displayed under Execute; the relevant system components to be tested are
highlighted light blue in the Schematic View (BEP, Peripherals, Keyboard), as shown below:
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
3.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
Figure 7-80 Full Back End Test w/o Hard Disk Scan