GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-119
9) After reviewing the test results, proceed as follows:
If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE HEALTHCARE On-Line Center for assistance.
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
7-6-1-3-3 CPU Tests
NOTE: For a description of these tests, refer to the information in CPU Tests on page 7 - 15.
1.) In the Diagnostic Test window, trackball to CPU Full Test in the list of CPU tests available for selection
from the Diagnostic Test Tree (see Figure 7-84 below) and press Set. The Test name is displayed
under Execute; the relevant system components to be tested are highlighted light blue in the
Schematic View (BEP, Peripherals, Keyboard), as shown below:
Figure 7-84 CPU Full Test