7-116 Section 7-6 - Performing Back End Diagnostics on the System
4.) Trackball to the Start button and press Set.
The Full Back End Test w/o Hard Disk Scan commences. While the test proceeds, the name of the
test currently in progress is displayed in the Execute field and the current Loop # and specific test
are shown below. The messages Init Done and Please Wait are displayed beside the progress bar
in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
Green: All recommended tests for these components have been completed and no problems
were identified.
Red: Problems were identified in these components during the performance of the test.
Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Full Back End Test w/o Hard Disk Scan sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests and sub-tests passed or failed, as shown in the example in Figure 7-81.
Note: For the Full Back End Test w/o Hard Disk Scan, many tests are performed in the test
sequence; these are listed in the Tests window as shown in Figure 7-81. For each currently-
selected test (for example, CPU Full Test) the corresponding report is listed in Reports on
the right.
For details of each report, refer to the explanations provided in CPU Tests on page 7 - 119:
7.) Use the scroll bars and/or arrows to scroll through the list of Test results and review the reports, as
required.
8) When done, trackball to the Exit button and press Set to close the View Report window.