GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
7-134 Section 7-6 - Performing Back End Diagnostics on the System
7-6-1-3-10 Monitor Test
NOTE: For a description of the monitor tests, refer to Monitor Test on page 7 - 18.
1.) In the Diagnostic Test window, trackball to Monitor Test in the list of Interactive PC Doctor tests
available for selection from the Diagnostic Test Tree (see Figure 7-99 below) and press Set. The Test
name is displayed under Execute; the relevant system components to be tested are highlighted light
blue in the Schematic View (BEP, Peripherals, Keyboard), as shown below:
2.) Trackball to the Start button and press Set.
The interactive PC Dr for Windows Monitor Test screen is displayed at the top left of the display
screen, as shown in Figure 7-100.
Figure 7-99 Monitor Test