DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-163
6.) On the Vivid i n/ Vivid q N system, type the Response code in the Response field in the pane
on the right (refer to Figure 7-136).
The following screen (Figure 7-138) displays a message stating that the Reload Setup was
successful.
7.) Reboot the system, as explained in the User Manual:
* Vivid i n/ Vivid q Nsystem: see "Switching On/Off", Vivid i n/ Vivid q N User Manual,
R2424431-100 Rev. 1, page 79.
The software reload process begins automatically.
Figure 7-138 Invoke SW Reload - Confirmation Message