EasyManua.ls Logo

GE Vivid i n - Page 556

GE Vivid i n
740 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-28 Section 8-2 - External Component Replacement Procedures
securing clips on all four sides, using both hands to release and lift it away from the LCD display.
8-2-8-5 LCD Display Frame Installation Procedure
NOTE: When performing the following installation procedure steps, it may be helpful to review the appropriate
movie clips previously referenced in the LCD Display Frame Removal Procedure.
1) With the LCD Display in the fully open position, return the LCD display frame to the correct position
and using both hands, press the frame under the securing clips on all four sides to fasten it against
the LCD display. Refer to Figure 8-32 on page 8-28.
2) Return and tighten the four screws (previously removed).
3) With the LCD Display in the open (upright) position, return the screw cap to each of the
four screws that fasten the LCD frame in position, as shown in Figure 8-30 on page 8-27.
4.) Proceed to perform the following functionality tests:
- SafeLock Cart - Grounding Continuity on page 10 - 23 - only if the Vivid i n/ Vivid q N system
is currently mounted on the SafeLock cart
- SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
Figure 8-32 Removing the LCD Display Frame

Table of Contents

Related product manuals