GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-51
8-3-6-5 Fan Installation Procedure
NOTE: When performing the following installation procedure steps, it may be helpful to review the appropriate
movie clips previously referenced in the Fan Removal Procedure.
1) Connect the replacement Fan cable to the BEP- refer to Figure 8-61, above.
2) Using both hands, carefully mount the replacement Fan Assembly into position on the Bottom
Assembly - refer to Figure 8-60 on page 8-50.
3) Make sure the Fan Assembly is seated in the correct position. Return and fasten the two screws
(previously removed) to secure the Fan Assembly to the Bottom Assembly, as shown in
Figure 8-58 on page 8-49.
4.) Reconnect the speaker cable to the fan assembly, as shown in Figure 8-59.
5) Return the Control Panel and Keyboard, as described in the Control Panel and Keyboard
Installation Procedure on page 8 - 11.
6) Return the bearing handle, as described in the Bearing Handle Installation Procedure on page 8 - 4.
7.) Proceed to perform the following functionality tests:
- SafeLock Cart - Grounding Continuity on page 10 - 23 - only if the Vivid i n/ Vivid q N system
is currently mounted on the SafeLock cart
- SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
- Calibration Tests on page 7 - 25