8-50 Section 8-3 - Internal Component Replacement Procedures
5) Using both hands, lift the Fan Assembly upwards from the front and then the rear to remove it from
the Bottom Assembly; lay it down adjacent to the system, ready for the next step as shown in
Figure 8-60 below.
Figure 8-60 Removing the Fan Assembly
6) Disconnect the Fan cable from the BEP and completely remove the Fan Assembly from the system,
as shown in Figure 8-61 below.