GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-70 Section 8-4 - SafeLock Cart Components Replacement
4) Proceed to perform the following functionality tests:
- SafeLock Cart - Grounding Continuity on page 10 - 23
- SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
Figure 8-81 ECG Cable Protector Secured on SafeLock Cart
ECG Cable Protector
Mounted on
SafeLock Cart
ECG Cable Protector as seen when in Use
ECG Cable
Connected to Scanner