GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-101
7. Select option: 2.
Vivid_q Full Clean Install (System + Application
Software)
Wait until the following message is displayed, prompting you to select the
required option:
8. Select option A.
Clean "Vivid q" Install System S/W ...
The user is prompted for a password.
9. When prompted, type the following password:
geulsservice and press <Enter>.
Note: When typing the password, use lower-case
letters.
A message is displayed.
The DVD then begins the software installation process and a progress indicator
is displayed, as shown below:
This stage of the process takes approximately 5 mins.
Note: Do not stop the software installation operation during this process!
10. Press <CTRL+ALT+DELETE>. The Ghosting process continues for approximately 10 minutes. The system
automatically loads the application and the system is ready for software option
installation.
11. Install software options as described in Chapter 3,
section 3-9-5 on page 3-84.
Table 8-5 Step-by-Step Vivid i n/ Vivid q N Software Upgrade Procedure (cont’d)
Step Instruction Expected Result