DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-102 Section 8-5 - Software Loading
12. To verify the software version, press Config and click
About at the bottom of the screen.
The About screen opens, and from the System Version tab, the installed
software version is displayed on the screen
13. Verify that the software version corresponds with the
System SW version you have installed (check against
the CD label on the installation disc) and that it was
correctly installed.
14. Restore all Presets as per user manual.
Note: When performing "Restore", click Admin at the
bottom of the screen, then click the Restore tab.
Select "System Configuration" and then de-select "GSP
Resources".
15. Install security patch as required.
16. Install peripherals.
Table 8-5 Step-by-Step Vivid i n/ Vivid q N Software Upgrade Procedure (cont’d)
Step Instruction Expected Result
De-select GSP Resources