EasyManua.ls Logo

GE Vivid i n - Page 635

GE Vivid i n
740 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-107
17. Navigate to the TCPIP tab and press the Set button on
the control panel.
Connectivity dialog box - TCPIP tab activated:
18. Navigate to the Save Settings button and press the Set
button on the control panel.
A message is displayed prompting for system re-boot:
19. Click OK and re-boot the system.
20. If this is a new software version, an adhesive Rating
Plate is supplied with the software CD.
Adhere the plate in the appropriate position under the
Bottom Assembly.
21. Proceed to perform the following functionality tests:
Software Configuration Checks on page 4 - 37
Calibration Tests on page 7 - 25
Full System Test on page 7 - 10
Table 8-6 Vivid i n/ Vivid q N Software Re-Installation (cont’d)
Step Instruction Expected Result

Table of Contents

Related product manuals