GE HEALTHCARE
DIRECTION FQ091013, REVISION 1    VIVID I N AND VIVID Q N SERVICE MANUAL
8-106 Section 8-5 - Software Loading
13. Press any key to begin the PMC upgrade. The PMC upgrade commences and the following screen with current PMC 
version and colored progress bar is displayed:
Note: If this screen fails to appear, refer to Installation of PMC Version on page 
8 - 135.
When done (after several minutes), the screen will turn blank and the fans will 
continue running. 
14. Power down the system by removing the AC adapter 
from the rear of the system.
Wait 10 seconds, then re-connect the battery and the 
AC adapter.
Boot-up the system by pressing the On/Off button.
If this is a new installation, the system will prompt for the software licence 
Password, as shown below. Proceed to step 15.
If not a new installation, there is no prompt for the software licence Password 
and the boot-up screen is displayed (see below). Proceed to step 16.
15. Type the password provided for the specific software 
and approve by clicking OK.
The system will continue to boot-up into Normal Scanning mode:
16. From the System Configuration window, click the 
Connectivity tab at the bottom of the window.
The Connectivity dialog box opens.
Table 8-6    Vivid i n/ Vivid q N Software Re-Installation (cont’d)
Step Instruction Expected Result