Chapter 8 - Replacement Procedures 8-105
9. When prompted, type the following password:
geulsservice and press <Enter>.
Note: when typing the password, use lower-case letters.
A message is displayed.
The CD then begins the software installation process and a progress indicator
is displayed, as shown below:
This stage of the process takes approximately 5 mins.
10. Press <CTRL+ALT+DELETE>. The system re-boots.
11. Follow the instructions on the screen, clicking OK
(or Cancel) when prompted.
The machine performs a number of reboots before completing the installation.
As the installation proceeds, the system displays various installation screens.
The installation process will take approximately 15 minutes - DO NOT
ATTEMPT TO STOP THE INSTALLATION BEFORE COMPLETION!
When done, the following PMC Firmware upgrade message is displayed:
12. Move cursor over the displayed window and press the
SET button.
The following message is displayed
Table 8-6 Vivid i n/ Vivid q N Software Re-Installation (cont’d)
Step Instruction Expected Result