GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-104 Section 8-5 - Software Loading
8-5-1-2 Vivid i n/ Vivid q N Software Re-Installation Procedure
NOTE: The following software re-installation procedure is based on the Vivid i n system. The procedure
is the same as for the Vivid q N system.
Table 8-6 Vivid i n/ Vivid q N Software Re-Installation
Step Instruction Expected Result
1. Remove the battery and remove all peripherals currently
connected to the system, with the following exceptions:
Make sure no peripherals are connected to the USB
connectors except the DVD and the PCMCIA.
2. Press the <Eject> button of the DVD\CD-RW drive and
remove any media if present.
3. Insert the CD provided with the software kit into the
DVD\CD-RW.
The DVD\CD-RW drive loads the software CD.
4. Hold down the <ON/OFF> button for 5 seconds to
shutdown the system.
The shutdown options are displayed on-screen.
5. Select Full Shutdown. The system powers down.
6. Hold down the <ON/OFF> button for three seconds. The system powers ON and when system boot-up from the CD is complete, the
following message is displayed prompting you to select the required option:
7. Select option 1.
Vivid_i Full Clean Install (System + Application
Software)
Follow the on-screen instructions, then wait until the following message is
displayed, prompting you to select the required option:
8. Select option A.
Clean "Vivid i" Install System S/W ...
The user is prompted for a password.