GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-146 Section 8-6 - Performing a Complete System Back-up
8-6-1-4 Running the Import Procedure
NOTE: This procedure is applicable when importing files from various media devices.
1.) Log on to the system as described in Logging on to the System on page 8 - 141.
2.) Insert the media (where the data in the previous section is stored), into the system.
3.) From the keyboard, press the Patient button; on the screen, click Patient List.
The Search Create Patient screen appears - Figure 8-139.
4.) Click the Import soft-key at the bottom of the screen - see Figure 8-145.
NOTE: If the Import button is not displayed, click More consecutively until the Import button is displayed.
NOTE: A warning is displayed (Figure 8-147) - to indicate that only the database is imported to the target
device.
Figure 8-146 Database Export Import Warning
5.) Click OK.
The Import screen appears - see Figure 8-147.
6.) Under "From" in the Import screen, select the required media source from the drop-down list (see
Figure 8-147).
NOTE: If an Import / Export Conflict screen displays (this could occur if two separate examinations (this could
occur if two separate examinations were saved under the same name), click OK and continue.
Figure 8-145 Soft-key Import Button
Figure 8-147 Import Screen

Table of Contents